Display unit
|
Test pressure
|
kPa,Mpa、bar,PSI
|
Leakage
|
Pa,Pa/s,mL/min
|
Mainframe
size
|
312(L)×277(W)×155(H)
|
Resolution accuracy
|
±1%~±5%F.S
|
Minimum display
|
1Pa
|
Display range
|
0~±1999.9Pa
|
Precision guarantee range
|
0~±2000Pa
|
Sensor range
|
0~±2000Pa
|
Test pressure range
|
50-500kPa
|
Pressure
resistance of sensor
|
13.8MPa
|
Leakage volume display
|
The number of integers is 3
bits, and the mL decimal number is 1.
Display sampling frequency 10HZ
|
Leaking
upper and lower limits
|
Less than 1999.9Pa
|
Parameter channel number
|
32group(0~31)
|
Time setting
|
0~999s
|
Hardware
|
32bitCPU
|
Power Supply
|
AC220V±5%,50HZ
|
Test pressure source
|
In general, dry clean air using 0.2~0.3MPa
higher than the test pressure.
|
ambient temperature
|
The temperature is 5~40 centigrade.
|
The preservation temperature is 0
~ 70 C
|
|
Humidity
|
80%RH, below, no condensation
|
Control the IO interface
|
Open circuit collector specification interface
and relay output
|
Note: the single side height of both sides screws is 2mm
5.4-inch touch screen display, Chinese operation interface is clear and easy to understand;
Differential pressure resolution, high precision;
Multi-channel parameters, the parameter channel selection and setting can be easily performed;
8 output control, 8 control inputs;
With manual test mode and automatic test mode;
With serial RS232 level output;
The remote control contacts are rich, can be connected to the PLC, and can also directly drive external devices (such as cylinders);
The output terminals can be manually adjusted to make external control more flexible;
The bubble-coating function, that is, when the soapy water is soaked in water to visually detect the bubbles, the instrument maintains a pressure-holding state;
The latest 10000 sets of test data can be saved, and the test results can be queried by date;
Leak detection result data, automatically stored;
Rich in data statistics and analysis functions;
With power-off memory protection;
Optional features: (The following features are subject to additional charges)
1. Use U disk to export data function; export data to EXCEL, easy to edit on the computer;
2. Can directly drive the printer or scanner